Concentricity Calibration ​

Overview ​

Concentricity calibration establishes a unified measurement reference across different magnifications. Through lens concentricity correction, measurement results remain accurate and consistent at all magnification levels.

Important

This method applies to EV Semi-Automatic Series and C Manual Series machines. SP Fully Automatic Series machines have a dedicated full magnification calibration procedure.

Prerequisites ​

  • Controller modification permission has been obtained from your equipment supplier

Steps ​

Step 1: Open Manufacturer Settings and Reset the Crosshair ​

  1. After obtaining controller modification permission, open the manufacturer settings
  2. Reset the crosshair in the manufacturer settings

Easson2D open manufacturer settings

Easson2D reset crosshair

Step 2: Confirm Magnification Calibration ​

Ensure that all zoom lens magnifications have been calibrated.

Easson2D confirm magnification calibration

Step 3: Measure the Reference Circle ​

Measure any reference circle with a diameter no larger than 1 mm.

Easson2D measure reference circle

Step 4: Enter Automatic Lens Calibration ​

  1. Click Calibration in the menu bar
  2. Select Auto Lens Calibration
  3. Select Square Calibration Target

Easson2D enter automatic lens calibration

Step 5: Select Reference Circle and Calibrate ​

  1. In the calibration interface, select the corresponding reference circle
  2. Click the Calibrate button
  3. The software will automatically complete the concentricity calibration

Easson2D select reference circle and calibrate

Verification ​

After calibration is complete, you can check the calibration status for each magnification in the relevant interface to confirm that concentricity calibration has taken effect.

Notes ​

  • All magnifications must be calibrated before performing concentricity calibration; otherwise, the results may be inaccurate
  • The reference circle diameter should be no larger than 1 mm
  • Exercise caution when modifying parameters in manufacturer settings — it is recommended to proceed under the guidance of a qualified professional or the supplier
  • After calibration, perform verification measurements at different magnifications