Probe Calibration
Before using the probe for formal measurement, the probe must be calibrated first.
Easson2D compensates for probe contact positions through the reference probe diameter. Although the probe itself has a nominal diameter, such as 3 mm, it is still recommended to calibrate using a standard with known dimensions during actual use, so that probe measurement results match the standard dimensions.
1. Open Probe Settings
Click the top menu:
Probe → Probe Settings
to access probe-related settings.
In the probe settings, find:
Reference Probe Diameter
Click it to enter the reference diameter of the probe currently in use.
2. Enter the Probe's Actual Diameter
First, enter an initial value based on the actual specifications of the current probe.
For example, when using a probe with a diameter of:
3 mm
first set the reference probe diameter to:
3.000 mm
This value serves as the initial value for subsequent calibration.
3. Measure a Standard
Prepare a standard with known dimensions.
For example, a standard dimension of:
50.000 mm
Use Easson2D's Horizontal Circle function to measure the standard with the probe, and observe the final measurement result.
4. Compare the Measured Value with the Standard Value
Compare the probe measurement result with the standard's known dimension.
For example:
Standard dimension:
50.000 mm
Actual measurement:
49.980 mm
This indicates that the current measurement result is smaller than the standard dimension.
In this case, return to the probe settings, appropriately decrease the reference probe diameter, then measure again.
5. Adjust the Reference Probe Diameter
Gradually adjust the reference probe diameter based on the measurement results.
If:
The measured value is smaller than the standard value
then appropriately:
Decrease the reference probe diameter
After the change, measure the same standard again using the horizontal circle function and observe the new measurement result.
Do not change the value by a large amount at once. It is recommended to fine-tune gradually so that the appropriate reference probe diameter is easier to find.
6. Repeat Measurement and Adjustment
Probe calibration usually requires repeating:
Measure the standard → check the result → adjust the reference probe diameter → measure again
After multiple adjustments, the horizontal circle measurement result gradually approaches the standard's known dimension.
For example:
Standard dimension: 50.000 mm
First measurement: 49.980 mm → adjust the reference probe diameter
Second measurement: 49.994 mm → continue fine-tuning
Third measurement: 50.000 mm
When the measurement result reaches the standard dimension or meets the current calibration requirements, the probe calibration is complete.
Calibration Workflow
The entire probe calibration process can be summarized as:
Probe → Probe Settings → Enter the reference probe diameter → Measure the standard with the horizontal circle → Compare the measured value with the standard value → Adjust the reference probe diameter → Measure again → Repeat adjustment until the result reaches the standard value
After Calibration
After the reference probe diameter is calibrated, keep using the current calibrated probe settings for subsequent measurements.
You can then use the probe for measurements such as points, lines, circles, heights, flatness, and cylinders.